24 June 1998 Machine precision assessment in 3D/2D digital subtracted angiography image registration
Erwan Kerrien, Regis Vaillant, Laurent Launay, Marie-Odile Berger, Eric Maurincomme, Luc Picard
Author Affiliations +
Abstract
During an interventional neuroradiology exam, knowing the exact location of the catheter tip with respect to the patient can dramatically help the physician. An image registration between digital subtracted angiography (DSA) images and a volumic pre-operative image (magnetic resonance or computed tomography volumes) is a way to infer this important information. This mono-patient multimodality matching can be reduced to finding the projection matrix that transforms any voxel of the volume onto the DSA image plane. This modelization is unfortunately not valid in the case of distorted images, which is the case for DSA images. A classical angiography room can now generate 3D X-ray angiography volumes (3DXA). Since the DSA images are obtained with the same machine, it should be possible to deduce the projection matrix from the sensor data indicating the current machine position. We propose an interpolation scheme, associated to a pre-operative calibration of the machine that allows us to correct the distortions in the image at any position used during the exam with a precision of one pixel. Thereafter, we describe some calibration procedures and an associated model of the machine that can provide us with a projection matrix at any position of the machine. Thus, we obtain a machine-based 2D DSA/3DXA registration. The misregistration error can be limited to 2.5 mm if the patient is well centered within the system. This error is confirmed by a validation on a phantom of the vascular tree. This validation also yields that the residual error is a translation in the 3D space. As a consequence, the registration method presented in this paper can be used as an initial guess to an iterative refining algorithm.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Erwan Kerrien, Regis Vaillant, Laurent Launay, Marie-Odile Berger, Eric Maurincomme, and Luc Picard "Machine precision assessment in 3D/2D digital subtracted angiography image registration", Proc. SPIE 3338, Medical Imaging 1998: Image Processing, (24 June 1998); https://doi.org/10.1117/12.310906
Lens.org Logo
CITATIONS
Cited by 19 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Angiography

Calibration

Image registration

Sensors

Matrices

Cameras

Error analysis

RELATED CONTENT

Venµs geometric image quality commissioning
Proceedings of SPIE (September 25 2018)
Accuracy in image measure
Proceedings of SPIE (October 06 1994)
New methods of measuring and calibrating robots
Proceedings of SPIE (October 03 1995)
Geo-registration of aerial images by feature matching
Proceedings of SPIE (January 19 2006)

Back to Top