Perspectives of brain imaging with PET systems

  • Stan Majewski EMAIL logo

Abstract

In this partial review and partial attempt at vision of what may be the future of dedicated brain PET scanners, the key implementations of the PET technique, we postulate that we are still on a development path and there is still a lot to be done in order to develop optimal brain imagers. Optimized for particular imaging tasks and protocols, and also mobile, that can be used outside the PET center, in addition to the expected improvements in sensitivity and resolution. For this multi-application concept to be more practical, flexible, adaptable designs are preferred. This task is greatly facilitated by the improved TOF performance that allows for more open, adjustable, limited angular coverage geometries without creating image artifacts. As achieving uniform very high resolution in the whole body is not practical due to technological limits and high costs, hybrid systems using a moderate-resolution total body scanner (such as J-PET) combined with a very high performing brain imager could be a very attractive approach. As well, as using magnification inserts in the total body or long-axial length imagers to visualize selected targets with higher resolution. In addition, multigamma imagers combining PET with Compton imaging should be developed to enable multitracer imaging.


Corresponding author: Stan Majewski, Department of Biomedical Engineering, University of California, Davis, CA, USA, E-mail:

Acknowledgments

The author is thankful to Professors Pawel Moskalik and Ewa Stępień for their invitation to contribute to this special edition of the journal to cover the PET brain imaging but the author wanted also to emphasize the common aspects of the next-generation PET imagers, that bring to focus the necessary pragmatic approaches like the J-PET concept.

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

Addendum 1: Limits of spatial resolution in PET

The basic idea of the next-generation PET systems is to take the spatial resolution of the system to its limit imposed by the physics of the PET process, the motion correction accuracy, and the statistical event limitations in creating high-resolution images composed of many small 3D pixels with high enough event statistics in each pixel to produce “good quality” images. The key physical contributing factors are: positron range and a-collinearity between the emission directions of the two 511 keV annihilation photons (Figure 22).

Figure 22: Left and center: The positron range for several positron labels. Right: a-collinearity between the two 511 keV annihilation photons emitted back-to-back introduces error in position definition of the annihilation point, that increases with the diameter/size of the detector.
Figure 22:

Left and center: The positron range for several positron labels. Right: a-collinearity between the two 511 keV annihilation photons emitted back-to-back introduces error in position definition of the annihilation point, that increases with the diameter/size of the detector.

In addition, one needs to minimize the detector contribution by its proper design. In technical terms, it means to implement: (1) smallest possible physical size of the system, limiting the maximal distances between the opposed coincident radiation sensors detecting the two 511 keV annihilation photons from the coincidence pairs, and (2) high intrinsic spatial resolution of the detector modules, including the depth-of-interaction (DOI) measurement. And DOI resolution is especially important in compact systems.

Using the approximate theoretical formulas one can get an estimate that sub-mm resolution for a brain-sized or breast-sized imagers is possible. In addition, there is experimental evidence that implementing the above optimization strategy works in several successful implementations in small animal PET scanners, such as PawPET and few other [examples: 132], [133], [134]. In these implementations, still not yet fully perfected, the reconstructed spatial resolution of 0.5–0.7 mm FWHM was achieved using F-18-based imaging agents.

Many papers dealt in the past with spatial resolution limits of the PET detectors [Example: 135]. An example of the empirical formula for the spatial resolution limit for a pixelated PET detector is:

FWHM=1.25(d/2)2+b2(0.0022D)2+r2+p,

where d is width of the pixel, D is the distance between the opposed coincidence detector modules in the contribution term expressing the a-collinearity of the two coincident annihilation photons, r is the positron range, and b is the blurring effect due to motion and the last term is due to parallax effect (p). 1.25 is the empirical factor related to the degradation of the point spread function (PSF) due to the non-uniform sampling of the LOR in the FOV and the image reconstruction process. This value is estimated assuming an analytical reconstruction algorithm such as the filtered back projection (FBP). From the formula, there are several key contributions to the reconstructed spatial resolution: physical processes (positron range and a-collinearity of the two annihilation photons), detector defined pixelation and DOI resolution, and also motion-induced error.

From the formula, an improvement can be achieved by making a more compact detector, using smaller-sized detector pixels, lowering the parallax effect, and avoiding blurring due to the motion. As formula provides, error due to the a-collinearity of the two annihilation photons increases with the distance between the modules. Therefore, closer structures are preferred, also from the sensitivity point of view, as well as due to smaller size, lower weight and reduced cost.

Below are shown two examples of the predictions from the above formula:

  1. for a 25 cm diameter (tight brain size) detector and F18, 2 mm transversal detector resolution pixel size, and assuming no blurring due to motion or parallax effects, the resolution limit is 1.75 mm FWHM.

  2. as in (1) but for 1 mm intrinsic detector resolution pixel this limit value changes to 1.19 mm FWHM.

In the paper by Shibuya et all, the authors calculated the non-collinearity contribution only: ”For example, the limit of PET spatial resolution are calculated to be 0.5 mm for a 10-cm diameter scanner, 1.2 mm for a 40-cm diameter scanner, and 2.1 mm for an 80-cm diameter scanner.” These results are consistent with the formula we use above. To achieve better resolution than formula predicts, modeling of the positron range and of the a-collinearity is expected to provide some improvement.

DOI contribution. The need for good DOI resolution is illustrated in the plots below showing calculated radial spatial resolution for several LSO crystal thickness designs (source: Johan Nuyts, Lowen, Belgium) (Figure 23).

Figure 23: The need for good DOI resolution is illustrated in these plots showing calculated radial spatial resolution for several DOI values (private communication: Johan Nuyts, Loewen, Belgium). In this compact brain scanner example, the inner imager diameter of 25 cm and 3 mm (left plot) and 1.5 mm (right plot) LYSO scintillation pixel sizes were assumed. The DOI resolution dominates the performance of the scanner. To maintain ∼2–3 mm resolution in the whole brain volume in such a tight detector structure, DOI resolution should be in 2–3 mm range.
Figure 23:

The need for good DOI resolution is illustrated in these plots showing calculated radial spatial resolution for several DOI values (private communication: Johan Nuyts, Loewen, Belgium). In this compact brain scanner example, the inner imager diameter of 25 cm and 3 mm (left plot) and 1.5 mm (right plot) LYSO scintillation pixel sizes were assumed. The DOI resolution dominates the performance of the scanner. To maintain ∼2–3 mm resolution in the whole brain volume in such a tight detector structure, DOI resolution should be in 2–3 mm range.

In this compact brain scanner example, the inner diameter of 25 cm and 3 mm and 1.5 mm scintillation pixel sizes were assumed. DOI resolution in the 1–7 mm range was assumed. To maintain ∼1.5–2.00 mm resolution in the whole brain volume in such a tight detector structure, DOI resolution should be 2–3 mm. Many studies investigated how to achieve good DOI resolution [examples: 136], [137], [138].

In addition, it is important to take into account other contributing practical limiting to the usable spatial resolution, that are often underestimated, and are related to the brain position recording and motion correction through the duration of the scan, and the ability to apply these accurate time sensitive corrections to the recorded data stream, before the image reconstruction process. Patient comfort is an important contributing factor, especially during long scans during dynamic brain imaging sessions. It is critical in order for the high-resolution imaging protocols to be successful, to develop a robust, concurrent or complementary and yet partially overlapping motion correction system that can be used in different imaging situations. There is a lot of experimental evidence that excellent intrinsic spatial resolution cannot be realized as readily in clinical environment, unlike in the controlled phantom studies, if there is inadequate human motion correction.

A very important and often forgotten element of the spatial resolution discussion is the impact of the statistical power of the detected signal, limited by the injected doses, biological uptake of the particular imaging agent, sensitivity of the system, and timing of the imaging scan (when done and for how long). To benefit from the ∼1 mm range spatial resolution there has to be enough detected events during the scan to fill all the small 3D voxels in the reconstructed volumes and per each time bin (if dynamic analysis is performed). Insufficient number of events in the image voxels results in very noisy images, that cannot be utilized unless software filtering algorithms that add controlled image blurring, which is in fact negating the advantages of the intrinsically high spatial resolution. This argument can be also formulated as that achievable event statistics in fact sets the limit on the useful spatial resolution for the imaging tasks, also depending on the imaging agent.

High uptake of the specific imaging agents justifies the push for the best achievable detector performance, even if in the majority of cases the very high resolution will not be usable and the filtering of data in the processing phase, as mentioned above, will be equivalent to using system with moderate spatial resolution to start with.

However, as mentioned earlier, the practically achievable spatial resolution using the intrinsically high-resolution PET scanner, either in the clinic or even in the research center where better precautions can be taken, may still ultimately depend on how accurate is the motion correction of human subjects. Many novel approaches are investigated, with optical markerless techniques and using directly the PET scanner data leading, with sub-mm results reported [90], [91], [92], [93], [94], [95].

Addendum 2: The TOF advantage

PET is converting into TOF PET to benefit from the increased sensitivity and resolution. Many papers describe the progress and discuss the future [10], [11], [12], [13], [14], [15], [16]. Just very recently, a paper was published showing first reconstructionless images obtained at 32 ps FWHM [17].

The variance reduction of a TOF-PET system over a non-TOF PET system can be computed analytically for the center of a uniform cylinder and, assuming D >> Δx, and it equals:

VARnonTOFVARTOF=21n 22aDΔx, where D is the diameter of the cylinder and is the TOF spatial resolution. Therefore, the variance gain of one TOF-PET system over another is inversely proportional to the ratio of their TOF-resolutions. There is some confusion of the advantage that TOF offers due to the two different measures used to compare systems with different TOF performance. One measure is the “sensitivity”, proportional to the above-defined variance gain, and the second is the Signal-to-Noise ratio (SNR) gain, expressed as square root of the variance gain. The SNR measure is a more accurate description of the practical situation in detecting image features against the noise background in the reconstructed tomographic images. For comparing scanners, SNR is thought to be a more appropriate measure. Between 200 and 400 ps, the sensitivity gain is factor 2, and SNR gain is \sqrt(2) = 1.4. There is also an additional issue how SNR is defined. Usually for detection, SNR is defined as the ratio between contrast and background noise. While contrast is important for detection, it is more related to the tracer. Due to relatively small object size, in brain imaging, the advantage of TOF is not that large, and as discussed before, it comes at a high expense and may actually cause a net sensitivity decrease due to creation of additional mechanical issues, and resultant cracks in angular coverage.

References

1. Wahl, RL. Principles and practice of PET and PET/CT, 2nd ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2008.Search in Google Scholar

2. Lewellen, TK. Recent developments in PET detector technology. Phys Med Biol 2008;53:R287–317. https://doi.org/10.1088/0031-9155/53/17/r01.Search in Google Scholar

3. Lewellen, TK. The challenge of detector designs for PET. AJR 2010;195:301–9. https://doi.org/10.2214/ajr.10.4741.Search in Google Scholar

4. Peng, H, Levin, CS. Recent developments in PET instrumentation. Curr Pharmaceut Biotechnol 2010;11:555–71. https://doi.org/10.2174/138920110792246555.Search in Google Scholar

5. Lee, JS. Technical advances in current PET and hybrid imaging systems. Open Nucl Med J 2010;2:192–208. https://doi.org/10.2174/1876388x01002010192.Search in Google Scholar

6. Vaquero, JJ, Kinahan, P. Positron emission tomography: current challenges and opportunities for technological advances in clinical and preclinical imaging systems. Annu Rev Biomed Eng 2015:385–414. https://doi.org/10.1146/annurev-bioeng-071114-040723.Search in Google Scholar

7. Guerra, D, Belcari, N, Bisogni, M. Positron emission tomography: its 65 years. Riv Nuovo Cimento 2016;39. https://doi.org/10.1393/ncr/i2016-10122-6.Search in Google Scholar

8. Walrand, S, Hesse, M, Jamar, F. Update on novel trends in PET/CT technology and its clinical applications. Br J Radiol 2016;89.10.1259/bjr.20160534Search in Google Scholar PubMed PubMed Central

9. Jones, T, Townsend, D. History and future technical innovation in positron emission tomography. J Med Imag 2017;4. https://doi.org/10.1117/1.JMI.4.1.011013.Search in Google Scholar

10. Berg, E, Cherry, SR. Innovations in instrumentation for positron emission tomography. Semin Nucl Med 2018;48:311–31. https://doi.org/10.1053/j.semnuclmed.2018.02.006.Search in Google Scholar

11. Been Han, Y, Gyu Kang, H, Hyun Song, S, Bae Ko, G, Sung Lee, J, Jong Hong, S. SiPM-based dual-ended-readout DOI-TOF PET module based on mean-time method. J Instrum 2019;14. https://doi.org/10.1088/1748-0221/14/02/p02023.Search in Google Scholar

12. Surti, S. Update on time-of-flight PET imaging. J Nucl Med 2015;56:98–105. https://doi.org/10.2967/jnumed.114.145029.Search in Google Scholar

13. Surti, S, Karp, JS. Advances in time-of-flight PET. Phys Med 2016;32:12–22. https://doi.org/10.1016/j.ejmp.2015.12.007.Search in Google Scholar

14. Cates, JW, Levin, CS. Advances in coincidence time resolution for PET. Phys Med Biol 2016;61:2255. https://doi.org/10.1088/0031-9155/61/6/2255.Search in Google Scholar

15. Vandenberghe, S, Mikhaylova, E, D’Hoe, E, Mollet, P, Karp, JS. Recent developments in time-of-flight PET. EJNMMI Phys 2016;3:3. https://doi.org/10.1186/s40658-016-0138-3.Search in Google Scholar

16. Seifert, S, Schaart, DR. Improving the time resolution of TOF-PET detectors by double-sided readout. IEEE Trans Nucl Sci 2015;62:3–11. https://doi.org/10.1109/tns.2014.2368932.Search in Google Scholar

17. Kwon, SI, Roncali, E, Gola, A, Paternoster, G, Piemonte, C, Cherry, SR. Dual-ended readout of bismuth germanate to improve timing resolution in time-of-flight PET. Phys Med Biol 2019;64:105007. https://doi.org/10.1088/1361-6560/ab18da.Search in Google Scholar

18. Kwon, SI, Ota, R, Berg, E, Hashimoto, F, Nakajima, K, Ogawa, I, et al.. Ultrafast timing enables reconstruction-free positron emission imaging. Nat Photonics 2021;15:914–8. https://doi.org/10.1038/s41566-021-00871-2.Search in Google Scholar

19. Phelps, ME. PET: the merging of biology and imaging into molecular imaging. J Nucl Med 2000;41:661–81.Search in Google Scholar

20. Carson, RE, Kuo, PH. Brain-dedicated emissio tomography systems: a perspective on requirements for clinical research and clinical needs in brain imaging. IEEE Trans Radiat Plasma Med Sci 2019;3:254–61. https://doi.org/10.1109/trpms.2019.2912129.Search in Google Scholar

21. Hooker, JM, Carson, RE. Human positron emission tomography neuroimaging. Annu Rev Biomed Eng 2019;21:551–81. https://doi.org/10.1146/annurev-bioeng-062117-121056.Search in Google Scholar

22. Laruelle, M. Imaging synaptic neurotransmission with in vivo binding competition techniques: a critical review. J Cereb Blood Flow Metab 2000;20:423–51. https://doi.org/10.1097/00004647-200003000-00001.Search in Google Scholar

23. Fung, EK, Carson, RE. Cerebral blood flow with [15O]water PET studies using an image-derived input function and MR-defined carotid centerlines. Phys Med Biol 2013;58:1903–23. https://doi.org/10.1088/0031-9155/58/6/1903.Search in Google Scholar

24. Faul, M, Coronado, V. Epidemiology of traumatic brain injury. In: Handbook of clinical neurology, vol 127. Elsevier; 2015: 3–13 pp. https://doi.org/10.1016/b978-0-444-52892-6.00001-5.Search in Google Scholar

25. Morbelli, S, Garibotto, V, Van De Giessen, E, Arbizu, J, Chetelat, G, Drezgza, A, et al.. A Cochrane review on brain [18F]FDG PET in dementia: limitations and future perspectives. Eur J Nucl Med Mol Imaging 2015;42:1487–91. https://doi.org/10.1007/s00259-015-3098-2.Search in Google Scholar

26. Finnema, SJ, Nabulsi, NB, Eid, T, Detyniecki, K, Lin, SF, Chen, MK, et al.. Imaging synaptic density in the living human brain. Sci Transl Med 2016;8:348ra96. https://doi.org/10.1126/scitranslmed.aaf6667.Search in Google Scholar

27. Kelley, P, Evans, MDR, Kelley, J. Making memories: why time matters. Front Hum Neurosci 2018;12:400. https://doi.org/10.3389/fnhum.2018.00400.Search in Google Scholar

28. Chen, MK, Mecca, AP, Naganawa, M, Finnema, SJ, Toyonaga, T, Lin, SF, et al.. Assessing synaptic density in Alzheimer disease with synaptic vesicle glycoprotein 2A positron emission tomographic imaging. JAMA Neurol 2018;75:1215–24. https://doi.org/10.1001/jamaneurol.2018.1836.Search in Google Scholar

29. Funck, T, Palomero-Gallagher, N, Omidyeganeh, M, Lepage, C, Toussaint, PJ, Khalili, N, et al.. Towards a gold standard for validation of quantification methods for PET neuroreceptor imaging. In: 29th international symposium on cerebral blood flow, metabolism and function; 2019: Journal of Cerebral Blood Flow & Metabolism.Search in Google Scholar

30. Molnar, Z, Clowry, GJ, Sestan, N, Alzu’bi, A, Bakken, T, Hevner, RF, et al.. New insights into the development of the human cerebral cortex. J Anat 2019;235:432–51. https://doi.org/10.1111/joa.13055 [Epub 2019/08/03].Search in Google Scholar

31. Surti, S, Karp, J. Impact of detector design on imaging performance of a long axial field-of-view, whole-body PET scanner. Phys Med Biol 2015;60:5343. https://doi.org/10.1088/0031-9155/60/13/5343.Search in Google Scholar

32. Zhang, XZ, Zhou, J, Cherry, SR, Badawi, RD, Qi, JY. Quantitative image reconstruction for total-body PET imaging using the 2-meter long EXPLORER scanner. Phys Med Biol 2017;62:2465–85. https://doi.org/10.1088/1361-6560/aa5e46.Search in Google Scholar

33. Zhang, X, Zhou, J, Cherry, SR, Badawi, RD, Qi, J. Quantitative image reconstruction for total-body PET imaging using the 2-meter long EXPLORER scanner. Phys Med Biol 2017;62:2465–85. https://doi.org/10.1088/1361-6560/aa5e46.Search in Google Scholar

34. Cherry, SR, Badawi, RD, Karp, JS, Moses, WW, Price, P, Jones, T. Total-body imaging: transforming the role of positron emission tomography. Sci Transl Med 2017;9:eaaf6169. https://doi.org/10.1126/scitranslmed.aaf6169.Search in Google Scholar

35. Cherry, SR, Jones, T, Karp, JS, Qi, J, Moses, WW, Badawi, RD. Total-body PET: maximizing sensitivity to create new opportunities for clinical research and patient care. J Nucl Med 2018;59:3–12. https://doi.org/10.2967/jnumed.116.184028.Search in Google Scholar

36. Leung, EK, Judenhofer, MS, Cherry, SR, Badawi, RD. Performance assessment of a software-based coincidence processor for the EXPLORER total-body PET scanner. Phys Med Biol 2018;63:18NT01. https://doi.org/10.1088/1361-6560/aadd3c.Search in Google Scholar

37. Berg, E, Zhang, X, Bec, J, Judenhofer, MS, Patel, B, Peng, Q, et al.. Development and evaluation of mini-EXPLORER: a long axial field-of-view PET scanner for nonhuman primate imaging. J Nucl Med 2018;59:993–8. https://doi.org/10.2967/jnumed.117.200519.Search in Google Scholar

38. Badawi, RD, Shi, H, Hu, P, Chen, S, Xu, T, Price, PM, et al.. First human imaging studies with the EXPLORER total-body PET scanner. J Nucl Med 2019;60:299–303. https://doi.org/10.2967/jnumed.119.226498.Search in Google Scholar

39. Zhang, X, Xie, Z, Berg, E, Judenhofer, MS, Liu, W, Xu, T, et al.. Total-body dynamic reconstruction and parametric imaging on the uEXPLORER. J Nucl Med 2019;61:285–91. https://doi.org/10.2967/jnumed.119.230565 [Epub 2019/07/16].Search in Google Scholar

40. Zhang, X, Xie, Z, Berg, E, Judenhofer, M, Liu, W, Lv, Y, et al.. Total-body parametric imaging using kernel and direct reconstruction on the uEXPLORER. J Nucl Med 2019;60(1 Suppl):456.Search in Google Scholar

41. Zhang, X, Cherry, S, Badawi, R, Qi, J. Total-body dynamic PET imaging with 100-ms temporal resolution. Montreal, QC, Canada: World Molecular Imaging Congress; 2019.Search in Google Scholar

42. Deng, Z, Hu, D, Ding, Y, Dong, Y. A comparison of image quality with uMI780 and the first total-body uEXPLORER scanner. J Nucl Med 2019;60(1 Suppl):381.Search in Google Scholar

43. Karp, JS, Vishwanath, V, Geagan, M, Muehllehner, G, Pantel, A, Parma, M, et al.. PennPET explorer: design and preliminary performance of a whole-body imager. J Nucl Med June 2019;21. https://doi.org/10.2967/jnumed.119.229997.Search in Google Scholar

44. Pantel, AR, Viswanath, V, Daube-Witherspoon, ME, Dubroff, JG, Muehllehner, G, Parma, MJ, et al.. Human imaging on a whole-body imager. J Nucl Med 2019. https://doi.org/10.2967/jnumed.119.231845.Search in Google Scholar

45. Viswanath, V, Daube-Witherspoon, M, Karp, J, Surti, S. Lesion detectability in long axial field of view TOF PET scanners. J Nucl Med 2019;60(1 Suppl):107.Search in Google Scholar

46. Lyu, Y, Lv, X, Liu, W, Judenhofer, MS, Zwingenberger, A, Wisner, ER, et al.. Mini EXPLORER II: a prototype high-sensitivity PET/CT scanner for companion animal whole body and human brain scanning. Phys Med Biol 2019;64. https://doi.org/10.1088/1361-6560/aafc6c.Search in Google Scholar

47. [D] Vandenberghe, S, Moskal, P, Karp, JS. State of the art in total body PET. EJNMMI Phys 2020;7:35. https://doi.org/10.1186/s40658-020-00290-2.Search in Google Scholar

48. Moliner, L, Rodríguez-Alvarez, MJ, Catret, JV, González, A, Ilisie, V, Benlloch, JM. NEMA performance evaluation of CareMiBrain dedicated brain PET and comparison with the whole body and dedicated brain PET systems. Sci Rep 2019:9;15484. https://doi.org/10.1038/s41598-019-51898-z.Search in Google Scholar

49. Tashima, H, Yamaya, T. Proposed helmet PET geometries with add-on detectors for high sensitivity brain imaging. Phys Med Biol 2016;61:7205–20. https://doi.org/10.1088/0031-9155/61/19/7205.Search in Google Scholar

50. Ahmed, AM, Tashima, H, Yoshida, E, Nishikido, F, Yamaya, T. Simulation study comparing the helmet-chin PET with a cylindrical PET of the same number of detectors. Phys Med Biol 2017;62:4541–50. https://doi.org/10.1088/1361-6560/aa685c.Search in Google Scholar

51. Ahmed, M, Tashima, H, Yoshida, E, Yamaya, T. Investigation of the optimal detector arrangement for the helmet-chin PET – a simulation study. Nucl Instrum Methods Phys Res A 2017;858:96–100. https://doi.org/10.1016/j.nima.2017.03.044.Search in Google Scholar

52. Gong, K, Majewski, S, Kinahan, PE, Harrison, RL, Elston, BF, Manjeshwar, R, et al.. Designing a compact high performance brain PET scanner-simulation study. Phys Med Biol 2016;61:3681–97. https://doi.org/10.1088/0031-9155/61/10/3681.Search in Google Scholar

53. Schmidtlein, CR, Turner, JN, Thompson, MO, Mandal, KC, Haggstrom, I, Zhang, J, et al.. Performance modeling of a wearable brain PET (BET). In: Proceedings SPIE. San Diego, USA: SPIE; 2016:9788 p.10.1117/12.2217020Search in Google Scholar

54. Schmidtlein, CR, Turner, JN, Thompson, MO, Mandal, KC, Haggstrom Zhang, J, Humm, JL, et al.. Initial performance studies of a wearable brain positron emission tomography camera based on autonomous thin-film digital Geiger avalanche photodiode arrays. J Med Imaging 2017;4:011003. https://doi.org/10.1117/1.JMI.4.1.011003.Search in Google Scholar

55. Moskal, P, Kowalski, P, Shopa, RY, Raczyński, L, Baran, J, Chug, N, et al.. Simulating NEMA characteristics of the modular total-body J-PET scanner—an economic total-body PET from plastic scintillators. Phys Med Biol 2021;66:175015. https://doi.org/10.1088/1361-6560/ac16bd.Search in Google Scholar

56. Tai, YC, Wu, H, Pal, D, O’Sullivan, JA. Virtual-pinhole PET. J Nucl Med 2008;49:471–9. https://doi.org/10.2967/jnumed.107.043034.Search in Google Scholar

57. Jiang, J, Samanta, S, Li, K, Siegel, SB, Mintzer, RA, Cho, S, et al.. Augmented whole-body scanning via magnifying PET. IEEE Trans Med Imaging 2020;39:3268–77. https://doi.org/10.1109/TMI.2019.2962623.Search in Google Scholar

58. Jiang, J, Li, K, Wang, Q, Puterbaugh, K, Young, JW, Siegel, SB, et al.. A second-generation virtual-pinhole PET device for enhancing contrast recovery and improving lesion detectability of a whole-body PET/CT scanner. Med Phys 2019;46:4165–76. https://doi.org/10.1002/mp.13724.Search in Google Scholar

59. Jiang, J, Samanta, S, Li, K, Hamdi, M, Siegel, SB, Mintzer, R, et al.. Augmented whole-body scanning via magnifying PET. J Nucl Med 2020;61(1 Suppl):309. https://doi.org/10.1109/TMI.2019.2962623.Search in Google Scholar

60. Zhou, J, Qi, J. Theoretical analysis and simulation study of a high-resolution zoom-in PET system. Phys Med Biol 2009;54:5193. https://doi.org/10.1088/0031-9155/54/17/008.Search in Google Scholar

61. Zhou, J, Qi, J. Adaptive imaging for lesion detection using a zoom-in PET system. IEEE Trans Med Imag 2010;30:119–30. https://doi.org/10.1109/TMI.2010.2064173.Search in Google Scholar

62. Carson, R, Berg, E, Ramsey, B, Cherry, S, Du, J, Tao, F, et al.. Design of the NeuroEXPLORER, a next-generation ultra-high performance human brain PET imager. J Nucl Med 2021;62(1 Suppl):1120.Search in Google Scholar

63. Soret, M, Bacharach, SL, Buvat, I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48:932–45. https://doi.org/10.2967/jnumed.106.035774.Search in Google Scholar

64. Guillette, N, Sarrhini, O, Lecomte, R, Bentourkia, M. Correction of partial volume effect in the projections in PET studies. Conference: IEEE Nuclear Science Symposium 2010:3541–3. https://doi.org/10.1109/NSSMIC.2010.5874467.Search in Google Scholar

65. Erlandsson, K, Buvat, I, Pretorius, PH, Thomas, BA, Hutton, BF. A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology. Phys Med Biol 2012;57:R119–59. https://doi.org/10.1088/0031-9155/57/21/r119.Search in Google Scholar

66. Cysouw, MCF, Golla, SVS, Frings, V, Smit, EF, Hoekstra, OS, Kramer, GM, et al.. Partial-volume correction in dynamic PET-CT: effect on tumor kinetic parameter estimation and validation of simplified metrics. EJNMMI Res 2019;9:12. https://doi.org/10.1186/s13550-019-0483-z.Search in Google Scholar

67. Yang, J, Hu, C, Guo, N, Dutta, J, Vaina, LM, Johnson, KA, et al.. Partial volume correction for PET quantification and its impact on brain network in Alzheimer’s disease. Sci Rep 2017;7:13035. https://doi.org/10.1038/s41598-017-13339-7.Search in Google Scholar

68. Wienhard, K, Schmand, M, Casey, M, Baker, K, Bao, J, Eriksson, L, et al.. The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph. IEEE Trans Nucl Sci 2002;49:104–10. https://doi.org/10.1109/tns.2002.998689.Search in Google Scholar

69. Eriksson, L, Wienhard, K, Eriksson, M, Casey, ME, Knoess, C, Bruckbaer, T, et al.. The ECAT HRRT: NEMA NEC evaluation of the HRRT system, the new high-resolution research tomograph. IEEE Trans Nucl Sci 2002;49:2085–8. https://doi.org/10.1109/tns.2002.803784.Search in Google Scholar

70. VanVelden, FH, Kloet, RW, van Berckel, BN, Buijs, FL, Luurtsema, G, Lammertsma, AA, et al.. HRRT versus HR+ human brain PET studies: an interscanner test–retest study. J Nucl Med 2009;50:693–702. https://doi.org/10.2967/jnumed.108.058628.Search in Google Scholar

71. Yamaya, T, Yoshida, E, Obi, T, Ito, H, Yoshikawa, K, Murayama, H. First human brain imaging by the jPET-D4 prototype with a pre-computed system matrix. IEEE Trans Nucl Sci 2008;55:2482–92. https://doi.org/10.1109/tns.2008.2003252.Search in Google Scholar

72. Z Wang, W Yu, S Xie. A dedicated PET system for human brain and head/neck imaging, published in: 2013 IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC), https://doi.org/10.1109/NSSMIC.2013.6829112.Search in Google Scholar

73. NeuroPET/CT PhotoDiagnostic systems. Available from: https://www.photodiagnostic.com/petct.Search in Google Scholar

74. Grogg, KS, Toole, T, Ouyang, J, Zhu, X, Normandin, M, Johnson, K, et al.. NEMA and clinical evaluation of a novel brain PET-CT scanner. J Nucl Med 2016;57:646–52. https://doi.org/10.2967/jnumed.115.159723.Search in Google Scholar

75. CerePET. Brain biosciences. Available from: https://www.linkedin.com/company/brain-biosciences-inc.Search in Google Scholar

76. Gómez Herrero, JA, Navarro García, J, Carlos, M, Vicente, A, José, L, Serra, P, et al.. Development of a new device for the early diagnosis of Alzheimer’s disease. Rev Biomechan. http://www.biomecanicamente.org/item/1159-rb65-caremibrain-english.html?tmpl=component&print=1.Search in Google Scholar

77. CareMiBrain imager, ONCOVISION, Valencia, Spain, https://www.oncovision.com.Search in Google Scholar

78. Gaudin, E, Toussaint, M, Thibaudeau, C, Fontaine, R, Normandin, M, Petibon, Y, et al.. Simulation studies of the SAVANT high resolution dedicated brain PET scanner using individually coupled APD detectors and DOI encoding. J Nucl Med 2019;60:531.Search in Google Scholar

79. NeuroLF, Positrigo. Available from: https://www.positrigo.com/.Search in Google Scholar

80. Bläckberg, L, Sanchez, D, Borghi, G, Ballabriga, R, Sajedi, S, Gómez, S, et al.. High sensitivity and high resolution dynamic brain-dedicated TOF-DOI PET scanner. In: 2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). 978-78-1-7281-7693-2/20. https://doi.org/10.1109/NSS/MIC42677.2020.9507837.Search in Google Scholar

81. Gaudin, E, Toussaint, M, Thibaudeau, C, Paille, M, Fontaine, R, Lecomte, R. Performance simulation of an ultra-high resolution brain PET scanner using 1.2 mm pixel detectors. IEEE Trans Radiat Plasma Med Sci 2019;3:334–42. https://doi.org/10.1109/trpms.2018.2877511.Search in Google Scholar

82. Organ specific PET, prescient imaging. Available from: http://prescient-imaging.com/.Search in Google Scholar

83. González, AJ, Majewski, S, Sánchez, F, Aussenhofer, S, Aguilar, A, Conde, P, et al.. The MINDView brain PET detector, feasibility study based on SiPM arrays. Nucl Instrum Methods Phys Res 2016;818:82–90. https://doi.org/10.1016/j.nima.2016.02.046.Search in Google Scholar

84. Benlloch, JM, González, AJ, Pani, R, Preziosi, E, Jackson, C, Murphy, J, et al.. The MINDVIEW project: first results. Eur Psychiatry 2018. https://doi.org/10.1016/j.eurpsy.2018.01.002.Search in Google Scholar

85. Lecoq, P. Pushing the limits in time-of-flight PET imaging. IEEE Trans Radiat Plasma Med Sci 2017;1. https://doi.org/10.1109/trpms.2017.2756674.Search in Google Scholar

86. Prior, J. The 10-ps TOF PET: clinical applications, presented at the FATA2019: FAst Timing Applications for nuclear physics and medical imaging, 3–5 September 2019, Accademia degli Zelanti e dei Dafnici. Catania, Italy: Acireale. https://agenda.infn.it/event/18991/timetable/.Search in Google Scholar

87. Gasper Razdevsek, P, Dolenec, R, Peter, K, Majewski, S, Studen, A, Korpar, S, et al.. Multi-panel limited angle PET system with 50 ps FWHM coincidence time resolution: a simulation study. accepted for publication in Trans Radiat Plasma Med Sci 2021:2021. https://doi.org/10.1109/TRPMS.2021.3115704.Search in Google Scholar

88. Harmon, ES, Thompson, MO, Ross Schmidtlein, C, Turner, JN, Krol, A. Towards 50 ps TOF-PET for brain imaging. Proc. SPIE 10953, Medical Imaging 2019: Biomedical Applications in Molecular, Structural, and Functional Imaging 2019:1095303. https://doi.org/10.1117/12.2515123.Search in Google Scholar

89. The Siemens Biograph vision PET/CT. https://www.siemens-healthineers.com/en-us/molecular-imaging/pet-ct/biograph-vision.Search in Google Scholar

90. van Sluis, J, de Jong, J, Schaar, J, Noordzij, W, van Snick, P, Dierckx, R, et al.. Performance characteristics of the digital Biograph Vision PET/CT system. J Nucl Med 2019. https://doi.org/10.2967/jnumed.118.215418.Search in Google Scholar

91. Kyme, AZ, Se, S, Meikle, SR, Fulton, RR. Markerless motion estimation for motion-compensated clinical brain imaging. Phys Med Biol 2018;63:105018. https://doi.org/10.1088/1361-6560/aabd48.Search in Google Scholar

92. Lu, Y, Gallezot, JD, Naganawa, M, Fontaine, K, Toyonaga, T, Ren, S, et al.. Data-driven motion detection and event-by-event correction for brain PET. J Nucl Med. 2018;59(1 Suppl).Search in Google Scholar

93. Lu, Y, Gallezot, JD, Naganawa, M, Ren, S, Fontaine, K, Wu, J, et al.. Data-driven voluntary body motion detection and non-rigid event-by-event correction for static and dynamic PET. Phys Med Biol 2019;64:065002. https://doi.org/10.1088/1361-6560/ab02c2.Search in Google Scholar

94. Ren, S, Lu, Y, Bertolli, O, Thielemans, K, Carson, RE. Event-by-event non-rigid data-driven PET respiratory motion correction methods: comparison of principal component analysis and centroid of distribution. Phys Med Biol 2019;64:165014. https://doi.org/10.1088/1361-6560/ab0bc9.Search in Google Scholar

95. Sun, C, Fontaine, K, Revilla, E, Toyonaga, T, Gallezot, JD, Mulnix, T, et al.. A data-driven quality control method for head motion tracking in PET. IEEE NSS/MIC. 2019.Search in Google Scholar

96. Hurley, S, Spangler-Bickell, M, Deller, T, Bradshaw, T, Jansen, F, McMillan, A. Data-driven rigid motion correction of PET brain images using list mode reconstruction. J Nucl Med 2019;60(1 Suppl):1358.Search in Google Scholar

97. Shi, H, Du, D, Xu, J, Peng, Q. Assessment of dedicated brain PET designs with different geometries. IEEE MIC; 2013.Search in Google Scholar

98. Shi, H, Du, D, Xu, J, Peng, Q. PMT based pentagonal and hexagonal detector module designs for convex polyhedron PET systems. IEEE MIC; 2013.Search in Google Scholar

99. Han, S, Dong, Du, Xu, JF, Su, Z, Peng, Q. Design study of dedicated brain PET with polyhedron geometry. Technol Health Care 2015;23:S615–23.10.3233/THC-151000Search in Google Scholar PubMed

100. Xu, JF, Huang, Q, Weng, F, Zan, Y, Chen, J, Xie, S, et al.. Progresses in designing a high-sensitivity dodecahedral PET for brain imaging, nuclear science symposium. In: Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD); 2016.10.1109/NSSMIC.2016.8069608Search in Google Scholar

101. Derenzo, SE, Choong, W-S, Moses, WW. Monte Carlo calculations of PET coincidence timing: single and double-ended readout. Phys Med Biol 2015;60:7309. https://doi.org/10.1088/0031-9155/60/18/7309.Search in Google Scholar

102. Derenzo, SE. Monte Carlo simulations of time-of-flight PET with double-ended readout: calibration, coincidence resolving times and statistical lower bounds. Phys Med Biol 2017;62:3828–58. https://doi.org/10.1088/1361-6560/aa6862.Search in Google Scholar

103. Philadelphia, SSS, Karp, JS. Limited angle tomography with time-of-flight PET; 2014. US Patent US 8,698,087 B2, Apr. 15.Search in Google Scholar

104. Cal-Gonzalez, J, Rausch, I, Lalith, K, Sundar, S, Lassen, ML, Muzik, O, et al.. Hybrid imaging: instrumentation and data processing. Front Phys 2018:6. Article 47.https://doi.org/10.3389/fphy.2018.00047. www.frontiersin.org.Search in Google Scholar

105. Hong, KJ, Ho Jung, YCJ, Kang, J, Hu, W, Lim, HK, Huh, Y, et al.. A prototype MR insertable brain PET using tileable GAPD arrays. Med Phys 2013;40. https://doi.org/10.1118/1.4793754.Search in Google Scholar

106. González, AJ, Conde, P, Hernández, L, Herrero, V, Moliner, L, Monzó, JM, et al.. Design of the PET-MR system for head imaging of the DREAM project. Nucl Instrum Methods 2013;702:94–7.10.1016/j.nima.2012.08.028Search in Google Scholar

107. Cho, G, Choi, Y, Sung Lee, J, An, HJ, Jung, JH, Park, HW, et al.. Preliminary evaluation of a brain PET insertable to MRI. EJNMMI Phys 2014;1(1 Suppl):A13. https://doi.org/10.1186/2197-7364-1-S1-A13. http://www.ejnmmiphys.com/content/1/S1/A13.Search in Google Scholar

108. Galve, P, Catana, C, Herraiz, JL, Udía, JM. GPU based fast and flexible iterative reconstructions of arbitrary and complex PET scanners: application to next generation dedicated brain scanners. In: M-03 – 2020 IEEE NSS-MIC conference.Search in Google Scholar

109. Catana, C. Development of dedicated brain PET imaging devices – recent advances and future perspectives. J Nucl Med 2019. https://doi.org/10.2967/jnumed.118.217901.Search in Google Scholar

110. Nakamoto, R, Nakamoto, Y, Ishimori, T, Fushimi, Y, Kido, A, Togashi, K. Comparison of PET/CT with sequential PET/MRI using an MR-compatible mobile PET system. J Nucl Med 2017;2. https://doi.org/10.2967/jnumed.117.197665.Search in Google Scholar

111. Vaska, P, Woody, CL, Schlyer, DJ, Shokouhi, S, Stoll, SP, Pratte, JF, et al.. RatCAP: miniaturized head-mounted PET for conscious rodent brain imaging. IEEE Trans Nucl Sci 2004;51:2718–22. https://doi.org/10.1109/tns.2004.835740.Search in Google Scholar

112. Majewski, S, Proffitt, J. 2011. Compact and mobile high resolution PET brain imager. US Patent. 7,884,331.Search in Google Scholar

113. Majewski, S, Proffitt, J, Brefczynski-Lewis, J, Stolin, A, Weisenberger, AG, Xi, W, et al.. A silicon photomultiplier based wearable brain imager. In: Proceedings of the IEEE Nuclear Science Symposium and Medical Imaging Conference Record, Valencia, Spain, 23–29 October 2011. 4030–4 pp.10.1109/NSSMIC.2011.6153766Search in Google Scholar

114. Bauer, CE, Brefczynski-Lewis, J, Marano, G, Mandich, M-B, Stolin, A, Martone, P, et al.. Concept of an upright wearable positron emission tomography imager in humans. Brain Behav 2016;00:1–10: e00530. https://doi.org/10.1002/brb3.530.Search in Google Scholar

115. Melroy, S, Bauer, C, McHugh, M, Carden, G, Stolin, A, Majewski, S, et al.. Development and design of next-generation head-mounted ambulatory microdose positron-emission tomography (AM-PET) system. Sensors 2017;17:1164. https://doi.org/10.3390/s17051164.Search in Google Scholar

116. Majewski, S and Brefczynski-Lewis, J. VIRPET – combination of virtual reality and PET brain imaging, US Patent 9,655,573, 2017.Search in Google Scholar

117. Noble, RM. Ambulatory microdose PET: a wearable PET scanner for neurologic imaging. J Nucl Med Technol 2019;47. https://doi.org/10.2967/jnmt.119.228718.Search in Google Scholar

118. Volkow, N. presentation at the BRAIN initiative workshop: transformative non-invasive imaging technologies. 2021. https://www.youtube.com/watch?v=A2p1oznwo6Y NIH Brain Initiative Transformative.Search in Google Scholar

119. Non-invasive imaging technologies workshop. 2021. https://videocast.nih.gov/watch=40183.Search in Google Scholar

120. Yoshida, E, Tashima, K, Nagatsu, K, Tsuji, AB, Kamada, K, Parodi, K, et al.. Whole gamma imaging: a new concept of PET combined with Compton imaging. Phys Med Biol 2020;65:125013. https://doi.org/10.1088/1361-6560/ab8e89.Search in Google Scholar

121. Yoshida, E, Tashima, H, Nagatsu, K, Tsuji, AB, Kamada, K, Parodi, K, et al.. Whole gamma imaging: a new concept of PET combined with Compton imaging. Phys Med Biol 2020;65:125013. https://doi.org/10.1088/1361-6560/ab8e89.Search in Google Scholar

122. Tashima, H, Yoshida, E, Wakizaka, H, Takahashi, M, Nagatsu, K, Tsuji, AB, et al.. Development of a hybrid image reconstruction algorithm combining PET and compton events for whole gamma imaging. In: 2020 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). https://doi.org/10.1109/NSS/MIC42677.2020.9507841.Search in Google Scholar

123. Moskal, P, Stępień, EŁ. Prospects and clinical perspectives of total-body PET imaging using plastic scintillators. Pet Clin 2020;15:439–52. https://doi.org/10.1016/j.cpet.2020.06.009.Search in Google Scholar

124. Moskal, P, Krawczyk, N, Hiesmayr, BC, Bala, M, Curceanu, C, Czerwiński, E, et al.. Feasibility studies of the polarization of photons beyond the optical wavelength regime with the J-PET detector. Eur Phys J C 2018;78:970. https://doi.org/10.1140/epjc/s10052-018-6461-1.Search in Google Scholar

125. Moskal, P, Kisielewska, D, Curceanu, C, Czerwińska, E, Dulski, K, Gajos, A, et al.. Feasibility study of the positronium imaging with the J-PET tomograph. Phys Med Biol 2019. https://doi.org/10.1088/1361-6560/aafe20.Search in Google Scholar

126. Moskal, P. Positronium imaging. In: 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) (IEEE, 2019). 1–3 pp. https://doi.org/10.1109/NSS/MIC42101.2019.9059856.Search in Google Scholar

127. Moskal, P, Jasińska, B, Stępień, EŁ, Bass, SD. Positronium in medicine and biology. Nat Rev Phys 2019;1:527–9. https://doi.org/10.1038/s42254-019-0078-7.Search in Google Scholar

128. Moskal, P, Dulski, K, Chug, N, Curceanu, C, Czerwiński, E, Dadgar, M, et al.. Positronium imaging with the novel multiphoton PET scanner. Sci Adv 2021;7. https://doi.org/10.1126/sciadv.abh4394.Search in Google Scholar

129. Moskal, P, Gajos, A, Mohammed, M, Chhokar, J, Chung, N, Curceanu, C, et al.. Testing CPT symmetry in ortho-positronium decays with positronium annihilation tomography. Nat Commun 2021;12:5658. https://doi.org/10.1038/s41467-021-25905-9.Search in Google Scholar

130. Stępień, E, Kubicz, E, Grudzień, G, Dulski, K, Leszczyński, B, Moskal, P. Positronium life-time as a new approach for cardiac masses imaging. Eur Heart J 2021;42(1 Suppl). https://doi.org/10.1093/eurheartj/ehab724.3279.Search in Google Scholar

131. Jasińska, B, Moskal, P. A new PET diagnostic indicator based on the ratio of 3gamma/2gamma positron annihilation. Acta Phys Pol B 2017;48:1737.10.5506/APhysPolB.48.1737Search in Google Scholar

132. Godinez, PF, Gong, K, Zhou, J, Judenhofer, MS, Chaudhari, AJ, Badawi, RD. Development of an ultra high resolution PET scanner for imaging rodent paws: PawPET. IEEE Trans Radiat Plasma Med Sci 2018;2. https://doi.org/10.1109/TRPMS.2017.2765486.Search in Google Scholar

133. Gaudin, E, Thibaudeau, C, Arpin, L, Fontaine, R, Lecomte, R. Imaging performance of a submillimetric spatial resolution APD-based preclinical PET scanner dedicated to mouse imaging. Atlanta: presented at the IEEE MIC; 2017:21–8 pp.Search in Google Scholar

134. Yamamoto, S, Watabe, H, Watabe, T, Ikeda, H, Kanai, Y, Ogata, Y, et al.. Development of ultrahigh resolution Si-PM-based PET system using 0.32 mm pixel scintillators. Nucl Instrum Methods Phys Res 2016;836:7–12. https://doi.org/10.1016/j.nima.2016.08.045.Search in Google Scholar

135. Moses, WW. Fundamental limits of spatial resolution in PET. Nucl Instrum Methods Phys Res 2011;648(Suppl):S236–40. https://doi.org/10.1016/j.nima.2010.11.092.Search in Google Scholar

136. Niknejad, T, Pizzichemi, M, Stringhini, G, Auffray, E, Bugalho, R, Silva, JCD, et al.. Development of high-resolution detector module with depth of interaction identification for positron emission tomography. Nucl Instrum Methods Phys Res Sect A Accel Spectrom Detect Assoc Equip 2017;845(C Suppl):684–8. https://doi.org/10.1016/j.nima.2016.04.080http://www.sciencedirect.com/science/article/pii/S0168900216302935.Search in Google Scholar

137. Du, J, Bai, X, Gola, A, Acerbi, F, Ferri, A, Piemonte, C, et al.. Performance of a high-resolution depth-encoding PET detector module using linearly-graded SiPM arrays. Phys Med Biol 2018;63:035035. https://doi.org/10.1088/1361-6560/aaa707.Search in Google Scholar

138. Du, J, Bai, X, Cherry, SR. Performance comparison of depth-encoding detectors based on dual-ended readout and different SiPMs for high-resolution PET applications. Phys Med Biol 2019;64:15NT03. https://doi.org/10.1088/1361-6560/ab1c37.Search in Google Scholar

Received: 2021-11-15
Accepted: 2021-11-17
Published Online: 2021-12-22

© 2021 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 26.4.2024 from https://www.degruyter.com/document/doi/10.1515/bams-2021-0178/html
Scroll to top button