ABSTRACT
One of the key parameters that determine the performance of the orthopaedic screw is the primary stability such as pull-out strength. The self-tapping flute designs have been adapted to the cortical screw to decrease insertion torque and increase pull-out strength, but the fluted tip resulted in decreasing of the screw strength. Therefore, this research was to study the different designs of the self-tapping screw. This study will use finite element analysis to determine the most appropriate screw design including the number and size of the flute, by the method of pull-out strength and bending strength analysis, before running the mechanical testing to validate the design and finite element analysis results. The analysis was based on the mechanical testing standard for metallic bone screw ASTM F543-17. The flute shape consists of non-fluted, edge fluted, bowl fluted and spiral fluted. The finite element analysis results show that the most appropriate screw designs are with 4 flutes with the size of 3 pitches from the tip, which overall, the bowl fluted is the optimal design for this self-tapping cortical screw which has the most appropriate pull-out strength and bending strength performance.
- ASTM International. 2019. ASTM F543 - 17: Standard Specification and Test Methods for Metallic Medical Bone Screws. American Society for Testing and Material InternationalGoogle Scholar
- Shu-wei Wu, Chia-Ching Lee, Ping-Yuen Fu, Shang-Chih Lin. 2011. The effects of flute shape and thread profile on the insertion torque and primary stability of dental implants. J. Medical Engineering & Physics, 34 (2012), 797-805. https://doi.org/10.1016/j.medengphy.2011.09.021Google Scholar
- Scott Yerby, Corey Scott, Nathan Evans, Katie Messing, Dennis Carter. 2001. Effect of Cutting Flute Design on Cortical Bone Screw Insertion torque and Pull-out strength. J. Orthop Trauma, Vol. 15, No. 3 (2001), 216-221. https://doi.org/10.1097/00005131-200103000-00012Google ScholarCross Ref
- ASTM International. 2020. ASTM F1839 - 08: Standard Specification for Rigid Polyurethan Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instrument. American Society for Testing and Material InternationalGoogle Scholar
- ASTM International. 2019. ASTM F136 - 13: Standard Specification for Wrought Titanium-6Aluminium-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Application (UNS R56401). American Society for Testing and Material InternationalGoogle Scholar
- MatWeb: MATERIAL PROPERTY DATA. Sawbones Solid Rigid Polyurethane Foam 30 PCF Density. Retrieved July 1, 2020 from http://www.matweb.com/search/datasheet.aspx?MatGUID=09d08bbdc67040f894e7e17b51b70d7aGoogle Scholar
Recommendations
Analysis and design of cold-formed dimpled steel columns using Finite Element techniques
Dimpled steel products are produced from the combination of an innovative dimpling process and a traditional forming process such as cold-roll forming or press-braking. The wider use of cold-formed dimpled steel members has promoted considerable ...
Fatigue strength in laser welding of the lap joint
Finite element analysis and fatigue test are performed to estimate the fatigue strength for the laser-welded lap joint. Concentrating on the application of design and manufacturing technology using the laser-welded lap joints, the residual stress and ...
Stress analysis of normal and osteoarthritic femur using finite element analysis
Stress distribution within the normal and osteoarthritic femur is studied using plane stress finite element analysis. During the progress of osteoarthritis of hip, the resultant load transferred from the acetabulum to the femur, instead of acting over a ...
Comments